By BioDockify Computational Research Team · November 21, 2026 · Tutorial
Build an HPLC calibration curve the ICH Q2 way: five levels, regression, residuals, back-calculated recovery, r-squared interpretation, and the acceptance limits that make the curve defensible — with a worked dataset.
Read the complete article with tables, code and references on BioDockify.
Key References
- ICH Harmonised Guideline Q2(R2): Validation of Analytical Procedures (adopted 2023). https://www.ich.org
- USP General Chapter <621> Chromatography. United States Pharmacopeia-National Formulary (current edition).
- USP General Chapter <1058> Analytical Instrument Qualification.
- ISO/IEC 17025:2017, General requirements for the competence of testing and calibration laboratories.
Scope & Limitations
A calibration curve is only as traceable as its standard — impure reference material biases every level identically and invisibly r² alone cannot detect curvature or heteroscedastic residuals; the plot is mandatory, not decorative Linear range established for one matrix and system does not transfer automatically to another